WS-ISOBOX - ISO BOX system for

ISO BOX system for temperature and humidity WINECAP™ wireless datalogger "on-field" calibration. Validated isothermal cell for calibration verify with comparative method using a provided referral TH datalogger ACCREDIA certified.

€ 1.185,80 TVA et frais de port exclus
Nous contacter
  • Données techniques
  • Téléchargements
Box dimensions 38x36x31cm
Internal cells dimensions 10x3cm - scalable depending on needs.
Maximum number of probes/datalogger can be calibrated (for each calibration session) 26 (WSD00T model case - 80x80x25mm)
Reference temperature and humidity and comparison Indoor temperature and relative humidity
Reference datalogger WSD02H-T1KM1
Included accessories Transportation strap
Assessment time Probes/dataloggers to be calibrated, must rest in the closed at least for 45 minutes before measurement starting.

WSD02H-T1KM1

Power supply 2.4A/h - 3.6V type “AA” lithium internal battery
Battery life (*) Up to 5 years (samples every 10 minutes and radio signal quality at least sufficient)
Measures acquired (2 channels)

· Indoor temperature
· Relative humidity

Sampling interval (*) Selectable from two minutes to 24 hours
Datalogger capacity 128,000 samples
Working temperature · Operative: -30°C ÷ +60°C
· Warehousing: -40°C ÷ +70°C
Radio frequency ISM 868MHz
Radio coverage (**) Up to 6Km on sight
Sealing IP30
Dimensions 80x80x25mm
Weight 100g
Case material ABS
Temperature - Transducer type Class A cabled PT1000 (20cm)
Temperature - Measure range 0°C ÷ +40°C
Temperature - Measure accuracy ±0.2°C - Range 0°C ÷ 40°C
Temperature - Measure resolution 0.01°C
Relative humidity - Transducer type CMOSens® Technology
Relative humidity - Measure range 0 ÷ 100%
Relative humidity - Measure accuracy ±2.5% from 0 to 90%From 90 to 100% change from ±2.5% to ±3.5%Accuracies are granted between 0°C ÷ 50°C range
Relative humidity - Measure resolution 0.05%RH

 * battery life and sampling interval may be influenced by fieldwork conditions and system configuration.

** radio coverage reachable using up to 32 WR12 repeaters (maximum 16 for each path) between the probe/datalogger and the gateway.