FR
EN IT FR

Accueil

Datalogger

WLRD40COPM - CO₂ concentration - Atmospheric pressure - PM2.5 particulate - PM10 particulate

WineCap™ LoRa® 4-channel Data Logger - CO2 concentration, atmospheric pressure, PM2.5 particulate matter and PM10 particulate matter.

€ 1.076,90 TVA et frais de port exclus
Nous contacter
  • Données techniques
  • Téléchargements
  • Produits connexes
Power supply 2 internal lithium “D” batteries, 19 Ah - 3.6 V (BAT3)Optional external power supply via USB Type-C input (5 Vdc 1 A)
Battery life (*) Up to 4 years (acquisition every 60 minutes and at least sufficient radio signal)
Measured quantities (4 channels) · CO₂ – Carbon dioxide concentration [ppm]
· Atmospheric pressure [mbar]
· PM2.5 – Fine particulate matter [μg/m³]
· PM10 – Inhalable particulate matter [μg/m³]
Sampling interval (*) Selectable from 1 minute to 24 hours (default: 60 minutes)
Datalogger capacity 65 000 measurements (per channel)
Operating temperature · Operation: -10°C to +60°C· Storage: -40°C to +70°C
Radio frequency ISM 868 MHz
Line-of-sight radio coverage (**) Up to 15 km (extendable using WLRE battery-powered enhancers)
Protection rating IP30
Dimensions (W x H x D) 155 × 120 × 50 mm
Weight 500 g
Enclosure material ABS
Mounting 3-point mounting
Connections Wireless, Mini USB 2.0
CO₂ concentration – Measurement principle Dual-wavelength NDIR sensor with integrated auto-calibration
CO₂ concentration – Measurement range 0–5 000 ppm
CO₂ concentration – Measurement resolution 1 ppm
CO₂ concentration – Measurement accuracy 0–5 000 ppm: < ±(50 ppm + 3% of reading)
CO₂ concentration – Compensations Compensated for temperature and atmospheric pressure
CO₂ concentration – Calibrations No calibration required. No need for ambient air zeroing periods
Atmospheric pressure – Measurement range 700–1 100 mbar
Atmospheric pressure – Measurement resolution 0.1 mbar
Atmospheric pressure – Accuracy at 25°C Typical: ±2 mbar (20–80% RH)
Atmospheric pressure – Temperature dependence ±0.015 mbar/K
Particulate matter PM2.5 and PM10 – Measurement principle Laser-based scattering
Particulate matter PM2.5 and PM10 – Measurement range 0–1 000 μg/m³
PM2.5 – Measurement accuracy range 0–100 μg/m³ ±5 [μg/m³ + 5 % of reading]
PM2.5 – Measurement accuracy range 100–1000 μg/m³ ±10 % of reading
PM10 – Measurement accuracy range 0–100 μg/m³ ±25 μg/m³
PM10 – Measurement accuracy range 100–1000 μg/m³ ±25 % of reading
Particulate matter PM2.5 and PM10 – Measurement resolution 0.1 μg/m³
Particulate matter PM2.5 and PM10 – Certifications MCERTS

* battery life may be influenced by fieldwork conditions, sampling/measuring interval and system configuration.** radio coverage can be extended using up to 32 enhancer WLRE (maximum 16 for each path) between the device and the gateway.